A Straightforward Technique for Determining Phosphorus in Foodstuffs Using High-Resolution Continuum Source Electrothermal Atomic Absorption Spectrometry (HR-CS ETAAS)
| Autoři | |
|---|---|
| Rok publikování | 2025 |
| Druh | Článek v odborném periodiku |
| Časopis / Zdroj | Analytical Letters |
| Fakulta / Pracoviště MU | |
| Citace | |
| www | https://www.tandfonline.com/doi/full/10.1080/00032719.2025.2527258 |
| Doi | https://doi.org/10.1080/00032719.2025.2527258 |
| Klíčová slova | Foodstuffs; high-resolution continuum source electrothermal atomic absorption spectrometry; HR-CS ETAAS; phosphorus; spectral interferences |
| Popis | This study presents a high-resolution continuum source electrothermal atomic absorption spectrometry (HR-CS ETAAS) method for the determination of total phosphorus in food and plant samples. The method, based on atomic absorption at the 213.618?nm spectral line, demonstrates adequate sensitivity with a sub-mg/L limit of detection and a wide calibration range. Under optimized atomization conditions and with an excess of Pd/Mg(NO3)2 modifier, spectral interferences are effectively suppressed, making the procedure robust and suitable for various matrices. The method’s performance and applicability were demonstrated through the analysis of five certified reference materials. |