Dual LC column characterization for mass spectrometry-based small molecule profiling of human plasma and serum

Investor logo
Investor logo

Warning

This publication doesn't include Institute of Computer Science. It includes Faculty of Science. Official publication website can be found on muni.cz.
Authors

TKALEC Žiga KOUDELKA Štěpán KLÁNOVÁ Jana PRICE Elliott James

Year of publication 2025
Type Article in Periodical
Magazine / Source Analytica Chimica Acta
MU Faculty or unit

Faculty of Science

Citation
web https://www.sciencedirect.com/science/article/pii/S0003267025003368?via%3Dihub
Doi http://dx.doi.org/10.1016/j.aca.2025.343942
Keywords Metabolomics; Exposome; Chemicals; Aqueous normal-phase; Non-targeted; Retention time prediction
Attached files
Description Background: Analyte annotation confidence in untargeted liquid chromatography mass-spectrometry (LC-MS) based chemical analysis can be enhanced by leveraging retention time information. For this, the chromatographic characteristics of the analytical system used should be well characterized. In this study, we measured 604 diverse chemical standards to characterize a dual LC setup consisting of pentabromobenzyl (PBr) and type-C silica hydride (SiH) columns operating in reversed-phase (RP) and aqueous normal-phase (ANP) mode, respectively. Results: ANP and RP separations individually retained 40 % and 64 % of standards in cLogP range from-6.60 to 8.67 and-3.34 to 12.95, respectively. Using both columns, the coverage increased to 79 % of standards with cLogP range from-6.60 to 12.95 (median cLogP = 1.63). Retention selectivity follows the number of basic nitrogen atoms in the molecule on SiH column and polarity (cLogP) on PBr column. Column repeatability and reproducibility were tested in triplicate using a chemically diverse subset of 108 standards. Repeatability of retention times, peak widths and peak areas was 0.3 %, 14 %, 4 % for SiH column and 0.2 %, 12 %, 4 % for PBr column. Similarly, reproducibility was 15 %, 34 %, 30 % for SiH column and 9 %, 18 % and 34 % for PBr column. Predictive RT models were developed based on experimental RT data, achieving R2 values of 0.92 and 0.96, with mean absolute errors of 0.29 min and 0.27 min for SiH and PBr columns, respectively. Significance: As proof of concept, 129 metabolites were annotated in pooled human serum and plasma by matching standard or predicted RT on one or both columns. The RT models and MS2 spectra of standards are openly available, facilitating uptake of this well-characterized chromatographic system to increase confidence in analyte annotation.
Related projects:

You are running an old browser version. We recommend updating your browser to its latest version.

More info